Original Paper
Abstract
Background: Arterial pressure-based cardiac output (APCO) is a less invasive method for estimating cardiac output without concerns about complications from the pulmonary artery catheter (PAC). However, inaccuracies of currently available APCO devices have been reported. Improvements to the algorithm by researchers are impossible, as only a subset of the algorithm has been released.
Objective: In this study, an open-source algorithm was developed and validated using a convolutional neural network and a transfer learning technique.
Methods: A retrospective study was performed using data from a prospective cohort registry of intraoperative bio-signal data from a university hospital. The convolutional neural network model was trained using the arterial pressure waveform as input and the stroke volume (SV) value as the output. The model parameters were pretrained using the SV values from a commercial APCO device (Vigileo or EV1000 with the FloTrac algorithm) and adjusted with a transfer learning technique using SV values from the PAC. The performance of the model was evaluated using absolute error for the PAC on the testing dataset from separate periods. Finally, we compared the performance of the deep learning model and the FloTrac with the SV values from the PAC.
Results: A total of 2057 surgical cases (1958 training and 99 testing cases) were used in the registry. In the deep learning model, the absolute errors of SV were 14.5 (SD 13.4) mL (10.2 [SD 8.4] mL in cardiac surgery and 17.4 [SD 15.3] mL in liver transplantation). Compared with FloTrac, the absolute errors of the deep learning model were significantly smaller (16.5 [SD 15.4] and 18.3 [SD 15.1], P<.001).
Conclusions: The deep learning–based APCO algorithm showed better performance than the commercial APCO device. Further improvement of the algorithm developed in this study may be helpful for estimating cardiac output accurately in clinical practice and optimizing high-risk patient care.
doi:10.2196/24762
Keywords
Introduction
Cardiac output (CO; L/min), the amount of blood pumped from the left ventricle per minute, is the main determinant of oxygen delivery to the body, including to the brain and vital organs, and is an important monitoring parameter during hemodynamic optimization. It is sometimes referred to as the stroke volume (SV; mL/beat), calculated by dividing the CO by the heart rate (HR; beats per minute). Particularly, in the perioperative phase, hemodynamic optimization is directly related to postoperative complications, which are the third leading cause of death worldwide [
]. Patients’ outcomes can potentially be improved by applying immediate treatment to maintain CO within 4-8 L/min or maintain SV within 60-100 mL/beat during major surgery [ ]. Optimization of CO is also essential for high-risk patients [ ]. Early interventions for hemodynamic control can significantly reduce mortality by more than 20% in high-risk patients [ ].The thermodilution method using a pulmonary artery catheter (PAC) has been regarded as a gold standard for measuring CO in clinical practice [
]. However, owing to its invasiveness, the risks associated with placement limit its use to only cardiac surgery, liver transplantations, and some critically ill patients. Instead, arterial pressure-based cardiac output (APCO) methods have been proposed as a less invasive method for estimating CO from the arterial pressure waveform without the risk of complications associated with a PAC [ , ]. These methods estimate systemic vascular resistance from arterial pressure waveform and general patient characteristics and predict the SV. As the arterial line is less invasive and usually inserted for continuous blood pressure monitoring, APCO devices such as the FloTrac (Edwards Lifesciences, Irvine, CA, United States) or LiDCO Rapid (LiDCO Ltd, London, UK) are widely used in perioperative CO management. However, inaccuracies of the commercially available APCO devices have been reported, especially, in sepsis or liver transplantation patients [ , ]. Improvements of the algorithm by researchers are also not possible, because only a subset of the algorithm has been openly released.Recent advances in machine learning techniques have led to many new approaches to solving clinical problems [
]. Deep learning techniques, such as convolutional neural networks (CNNs), have performed well in bio-signal analysis [ ]. In contrast, the shortage of clinical bio-signal data makes it difficult to train deep learning models properly [ , ]. Publicly available bio-signal datasets are still limited compared with medical imaging or structured datasets [ - ]. Furthermore, data with reduced clinical use, such as PAC-based CO, worsen this tendency. In this case, after training a model with a relatively common dataset, a transfer learning technique can be used to refine the model parameters with relatively rare data. Previous studies also reported performance gains from transfer learning with bio-signal data [ - ].In this study, a novel APCO algorithm was built using a transfer learning technique. The algorithm learned the commercial APCO algorithm and then was trained with less-common PAC data. In addition, several preprocessing techniques were proposed to analyze the arterial pressure waveforms for predicting CO. Finally, the deep learning model was validated using real-world bio-signal data, which was collected during a separate period than the training data and includes cardiac surgery and liver transplantation patients. This study hypothesized that a model developed using real-world clinical data, deep learning techniques, and transfer learning techniques can be more accurate than a commercial APCO device for estimating CO.
Methods
Study Approval
All data used in this study were obtained from the prospective registry of the vital signs for surgical patients at Seoul National University Hospital. The registry was approved by the Institutional Review Board of Seoul National University Hospital (H-1408-101-605) and registered at the clinical trial registration site (ClinicalTrials.gov, NCT02914444). This retrospective study was also approved by the Institutional Review Board (H-2007-015-1138). However, the need for written informed consent was waived because of the anonymity of the data.
Data Collection
The registry collected synchronous vital signs and bio-signal data from various medical devices using the Vital Recorder Program [
]. Among the cases in the registry, those from between August 2016 and September 2019 were used in this study. The cases collected in the last 8 months of the study period (February 2019 to September 2019) were used for the testing dataset. The remaining cases were used for training the model.During data collection, CO monitors were used according to the discretion of the anesthesiologist. CO values were collected at 2-second intervals from the serial port of APCO devices, such as the EV1000 clinical platform or the Vigileo system (Edwards Lifesciences, Irvine, CA, United States) with a fourth-generation FloTrac algorithm or a PAC-based device such as the Vigilance II (Edwards Lifesciences, Irvine, CA, United States). The arterial pressure waveform was recorded at 500 Hz from the analog output port of the TRAM module (GE Healthcare, Chicago, IL, United States), and the heart rate was recorded at 2-second intervals from the serial port of the Solar 8000 patient monitor (GE Healthcare, Chicago, IL, United States). General patient characteristics (ie, age, sex, weight, and height) were collected from electronic medical records.
The deep learning model requires massive amounts of data for good performance. However, with PAC-based CO monitoring data, such as the Vigilance II, it is difficult to retain massive amounts of data, which hinders the ability to develop of a good deep learning model using real-world databases. Hence, the model was pretrained using APCO data from the EV1000 or Vigileo, from which data are relatively easy to obtain. After that, we tuned the model using PAC data from the Vigilance II, from which data are hard to obtain. In total, 1572 cases of surgery were recorded with APCO monitoring devices for pretraining, 290 cases were recorded with PAC-based CO monitoring devices for tuning or testing, and 195 cases were recorded with both APCO and PAC-based CO monitoring devices for tuning or testing. Among the 2057 cases, 1958 cases (95.19%) that were operated on from August 2016 to January 2019 were used for pretraining or tuning, and the remaining 99 cases (4.81%) that were operated on since February 2019 were used for testing.
Data Preprocessing
A dataset of arterial pressure waveforms was preprocessed, including corresponding SV values. The arterial pressure waveforms were resampled from 500 Hz to 100 Hz and sliced into 20-second segments. Each pair of a 20-second segment and an SV was referred to as a “sample.” For preprocessing the samples, the following 4 steps were performed: (1) converting the output of PAC-based CO data to SV, (2) smoothing the APCO data, (3) delaying the PAC data, and (4) removing unsuitable samples.
The APCO monitoring device provides the SV value, because it estimates the amount of blood moving from arterial waves when a stroke occurs. In contrast, a PAC-based CO monitoring device emits the CO value, because it measures the temperature change by blood flow from the pulmonary artery catheter. Owing to the physiological differences between the 2 methods, CO values needed to be converted from the Vigilance II monitor to SV, using HR values (SV = CO/HR), to synchronize with APCO data.
In addition, because there were larger fluctuations in the APCO data than clinically expected CO changes, the APCO data were smoothed using a locally weighted scatterplot smoother (LOWESS) algorithm [
]. We used the hyperparameter, λ=0.03, of the LOWESS algorithm. If APCO data were not recorded for more than 200 seconds in a single case due to recording errors, the LOWESS algorithm was applied separately.In addition, the PAC-based SV value was delayed, to synchronize the time differences between arterial pressure waveforms and PAC data. There are several minutes of delay in the CO values using the Vigilance II in its “Trend” mode [
, ]. To determine the time lag of the CO values in the Trend modes, we used the device with the “STAT” mode in several cases, allowing both the CO values and the “CO stat” values to be transferred. Then, the CO values were compared with the CO stat values, and we obtained the minimum mean absolute difference with the delay of 2 minutes. Thus, PAC-based SV values were shifted to the time 2 minutes earlier than the recorded time. More detailed descriptions and examples of delaying PAC values are provided in .Finally, unsuitable samples were removed for robust deep learning model training. Samples with a blood pressure <25 mmHg or >250 mmHg and an SV <20 mL or >200 mL were removed from the dataset. After applying a beat-detection algorithm, we eliminated samples with an HR <30 beats/min or >180 beats/min, a pulse pressure <20 mm Hg, or with frequent (>50%) ventricular premature beats [
].Model Building
A CNN model was designed to learn the appropriate feature extraction from the 20-second segments of arterial pressure waveforms. The goal of the model was to estimate PAC-based CO values using the arterial pressure waveform and the patient’s demographics (ie, age, sex, weight, and height) (
). The model consists of 2 parts: feature extraction and regression. The feature extraction part of the model was composed of 2 successive pairs of convolution and batch-normalization layers, 15 inception modules, 2 nonlocal modules with pooling, and dropout layers [ - ]. The regression part of the model, which was composed of 3 fully connected, batch-normalization, and dropout layers, takes a concatenation of the extracted features from the feature extraction part and the patient’s demographic information as input and returns a predicted SV value. For all layers, a rectified linear unit was used as the activation function.Input samples were abstracted by 2 consecutive pairs of convolution and batch-normalization layers (dropout rate of 0.5), before feeding them into the inception modules. The inception modules consisted of 4 paths with 6 convolution layers and 1 pooling layer. The filter size of each path in the inception module was 32, and the outputs of the 4 paths were concatenated to 128. The detailed configurations of the inception module are illustrated in
B. Note that p represents the pooling rate of the average pooling layer, k indicates the kernel size of the convolution layer, and d is the dilation rate of the dilated convolution layer. After an odd-numbered inception module, an average pooling layer intervened to reduce the size of feature maps from the previous inception module. The shaded inception module block in A was repeated 6 times. Nonlocal modules were added just before the last 2 inception modules to consider the global covariance in each segment. The first fully connected layer was a flattening layer (size of 132) that took the concatenation of the average of feature maps by global average pooling (size of 128) and demographic information (size of 4). The number of neurons in the last 2 fully connected layers was 64 and 1, respectively. Immediately after the second fully connected layers, there was a batch-normalization layer to achieve robustness for cases in which a batch is biased to a specific SV range. The last fully connected layer consisted of a single neuron, which represents a float value of the SV. For all fully connected layers, a dropout of 0.5 was used. The output dimensions of each module or layer are described in . Note that dimensions are represented as BS, channel, and length, where BS indicates batch size. The dotted arrows indicate a dropout of 0.5.Model Training
Our model was trained using a transfer learning method with 2 steps: pretraining and tuning. During pretraining, SV values were used from EV1000 or Vigileo to find rough parameters of the model for analyzing arterial pressure waveforms. The input of pretraining was 20-second segments of arterial pressure waveform and patient demographic information, and the output was the predicted SV. After that, the parameters were tuned using PAC data. The input and output of tuning were the same as in pretraining; however, the parameters of the regressor part were initialized with the Xavier algorithm to be retrained with PAC data [
]. Gradient-descent optimizers, RAdam, and Lookahead were used to update parameters [ , ]. The batch size was 512, and the loss function was the root mean squared error. Note that the loss function was calculated based on the equation (Ŷ = predicted value; Y = ground truth; N = batch size):Root mean squared error = (∑N(Ŷ-Y)2/N)1/2 (1)
The model was tested every 200 steps using 30% of the training datasets to calculate the validation errors. The training was stopped when the validation errors no longer decreased after 50 times and then was restarted with a smaller learning rate (decay rate of 0.5). The training was performed using our own program (code available at [
]), written in Python using PyTorch 1.1.0 on a graphics processing unit server with two 10-core Intel Xeon central processing units and eight Nvidia GTX 1080Ti graphics processing units.Statistical Analysis
The statistics of patient demographics were described in the training and testing groups, comparing the heterogeneity of the variables. Note that training groups contained pretraining and tuning datasets, and the testing group contained the testing dataset. For continuous variables (ie, age, height, and weight), a Mann-Whitney U test was performed for comparisons after testing for normality using a Shapiro-Wilk test. For categorical variables (ie, sex), a Pearson chi-square test was conducted for comparisons between groups.
The performance of the deep learning model was validated using error, absolute error, percentage error, and absolute percentage error, using the testing dataset. Each metric was calculated based on the equation (Ŷ = predicted value; Y = PAC value; N = number of samples):
Error (mL) = ∑N (Ŷ-Y)/N (2)
Absolute error (mL) = ∑N |Ŷ-Y|/N (3)
Percentage error (%) = [∑N {(Ŷ-Y)/Y}/N] × 100 (4)
Absolute percentage error (%) = {∑N |(Ŷ-Y)/Y|/N} × 100 (5)
Efforts were made to validate the generalizability and substitutability of our model by comparing its performance with the FloTrac in two radically different patient groups: the patients who underwent cardiac surgery and the patients who underwent liver transplantation surgery. Among the test dataset, both the FloTrac and Vigilance II devices were used simultaneously in 16 cases of cardiac surgery and 40 cases of liver transplantation surgery. With these 56 cases, direct comparisons were performed between the deep learning model and FloTrac using a paired t test for overall cases and each subgroup.
Spearman correlation coefficients were calculated between the SV of the deep learning model and PAC, and between the SV of the EV1000 or Vigileo and PAC. Bland-Altman analysis was used to test the agreement of either the pair of the deep learning model and PAC-based SV or the pair of the FloTrac and PAC-based SV [
]. Bias was defined as the mean difference between SVs, and the upper and lower limits of agreement were defined as ±1.96 SDs of the bias. The trending ability of the deep learning model was examined using a 4-quadrant plot analysis [ ]. The concordance rate of the association for percentage changes in SV was calculated between our model or the FloTrac and PAC, with the exclusion of 10% of the changes [ ].All data are expressed as the mean (SD), median (interquartile range), or absolute numbers (%). P values <.05 were considered statistically significant. Statistical analyses were performed using Python Scipy 1.4.1.
Results
Data from 2057 surgical cases (1232 general, 59.89%; 636 thoracic, 30.92%; 159 urologic, 7.73%; 23 gynecologic, 1.12%; 6 otolaryngologic, 0.29%; and 1 plastic surgery, 0.05%) in the registry were extracted and preprocessed. Of these 2057 surgical cases, we used the data from 1958 cases (95.19%) for training and 99 cases (4.81%; 59 cardiac surgery, 60%; 40 liver transplantation, 40%) for testing. For transfer learning, of 2057 surgical cases, we used the data from 1572 cases (76.42%) for pretraining and the data from 386 cases (18.77%; 245 cardiac surgeries, 63.5%; 141 liver transplantations, 36.5%) for tuning. Demographic information of the patients was not different between the training and testing datasets, except, that the patients in the testing dataset were slightly older (
).Characteristic | Training dataset (n=1958) | Testing dataset (n=99) | Statistical test | |
Method used | P value | |||
Age, median (interquartile range), years | 61.2 (51.2-69.5) | 63.8 (57.4-71.9) | Mann-Whitney U test | .02 |
Sex, number male (%) | 1195 (61.03%) | 66 (67%) | Pearson chi-square test | .309 |
Height, median (interquartile range), years | 164.0 (157.3-170.0) | 163.4 (156.7-168.0) | Mann-Whitney U test | .231 |
Weight, median (interquartile range), kg | 63.1 (55.2-72.4) | 64.0 (57.5-72.1) | Mann-Whitney U test | .494 |
The absolute error of the deep learning model for the testing dataset was 14.5 (SD 13.4) mL (
). In the subgroup analysis, the absolute errors of the deep learning model were 10.2 (SD 8.4) mL for cardiac surgery and 17.4 (SD 15.3) for liver transplantation.Measure | Overall (n=99), mean (SD) | Cardiac surgery (n=59), mean (SD) | Liver transplantation (n=40), mean (SD) |
Error (mL) | –4.4 (19.2) | 2.3 (13.0) | –9.0 (21.3) |
Absolute error (mL) | 14.5 (13.4) | 10.2 (8.4) | 17.4 (15.3) |
Percentage error (%) | 0.4 (27.4) | 9.0 (26.8) | –5.5 (26.2) |
Absolute percentage error (%) | 20.5 (18.2) | 20.5 (19.4) | 20.4 (17.4) |
In the testing dataset, the data from 56 cases with both PAC and FloTrac (16 cardiac surgery, 29%; 40 liver transplantation, 71%) data were used to compare the performance of the deep learning model with that of the FloTrac. The absolute error of the deep learning model was significantly lower than that of the FloTrac (16.5 mL vs 18.3 mL, P<.001;
). In the subgroup analysis, the absolute errors of the deep learning model were lower than those of the FloTrac, in both cardiac surgery (11.1 mL vs 14.3 mL, P<.001) and liver transplantation (17.4 mL vs 19.0 mL, P<.001; ). The individual plots of the time course of the measured and predicted SVs in all 99 testing cases are available in .Measure | Deep learning model, mean (SD) | FloTrac, mean (SD) | Statistical test | ||
Count, n | P value | ||||
Error (mL) | |||||
Overall (n=56) | –7.9 (20.7) | –8.4 (22.1) | 158725 | <.001 | |
Cardiac surgery (n=16) | –1.7 (15.2) | 2.7 (18.3) | 65260 | <.001 | |
Liver transplantation (n=40) | –9.0 (21.3) | –10.3 (22.2) | 93465 | <.001 | |
Absolute error (mL) | |||||
Overall (n=56) | 16.5 (14.8) | 18.3 (15.1) | 158725 | <.001 | |
Cardiac surgery (n=16) | 11.1 (10.5) | 14.3 (11.8) | 65260 | <.001 | |
Liver transplantation (n=40) | 17.4 (15.3) | 19.0 (15.4) | 93465 | <.001 | |
Percentage error (%) | |||||
Overall (n=56) | –4.4 (26.9) | –5.6 (28.6) | 158725 | <.001 | |
Cardiac surgery (n=16) | 1.8 (29.9) | 9.5 (34.9) | 65260 | <.001 | |
Liver transplantation (n=40) | –5.5 (26.2) | –8.3 (26.5) | 93465 | <.001 | |
Absolute percentage error (%) | |||||
Overall (n=56) | 20.3 (18.3) | 22.5 (18.5) | 158725 | <.001 | |
Cardiac surgery (n=16) | 19.3 (22.9) | 25.4 (25.7) | 65260 | <.001 | |
Liver transplantation (n=40) | 20.4 (17.4) | 22.0 (16.9) | 93465 | <.001 |
The Spearman rho value of our deep learning model was 0.64 (P<.001), whereas the rho value of the FloTrac was 0.57 (P<.001;
). Bland-Altman analysis demonstrated that the lower and upper limits of agreements, respectively, were –48.5 (95% CI –48.7 to –48.3) mL and 32.7 (95% CI 32.5-33.0) mL for the deep learning model and –51.8 (95% CI –52.0 to –51.5) mL and 35.0 (95% CI 34.8-35.3) mL for the FloTrac. The 4-quadrant plot showed concordance rates of 53% for the deep learning model and 46% for FloTrac ( ). Mean differences were smaller in our deep learning model than in the commercial APCO device.Discussion
Principal Results
In this study, we built and evaluated an open-source deep learning–based APCO algorithm using a large set of prospectively collected registry data. The performance of the deep learning model was better than that of the FloTrac, the commercially available APCO algorithm.
A mathematical analysis of the association between arterial blood flow and pressure waveform has a long history of over 100 years [
]. A key factor in this flow-pressure association is the estimation of systemic vascular resistance (SVR), because the flow is determined by the pressure gradient and vascular resistance [ ]. However, since SVR changes with the patient's condition, the coefficient also needs to be updated in real-time [ - ]. An uncalibrated APCO algorithm, which automatically updates the coefficient using the patient’s general characteristics and arterial pressure waveforms, can be a convenient solution in clinical situations. However, currently available commercial uncalibrated APCO devices have been reported to work poorly in patients with vasodilatory states such as sepsis or liver transplantation [ , , , , ]. Our results showed that our deep learning–based APCO algorithm outperformed the FloTrac algorithm in both cardiac surgery and liver transplantation patients. However, both the deep learning–based and FloTrac algorithms showed a positive bias in cardiac surgery patients, who usually have low SVs, and a negative bias in liver transplantation patients, who usually have high SVs. This tendency to return to the average is also shown in C and may be a fundamental limitation of the APCO algorithm, in which SVR should be estimated only from the arterial pressure waveform and patient’s demographics. Otherwise, this tendency may occur because most of the data obtained from routine clinical practice are within the normal SV range.Clinical, Academic, and Technical Implications
The measurement of CO is essential for clinical hemodynamic optimization. As the deep learning model is more accurate than the commercial APCO device, it can help enhance patient management and improve final outcomes. For example, goal-directed fluid therapy or SV optimization can be performed using our model. However, further validation and implementation are required for clinical applications. Disclosing our dataset and model, researchers can improve the model and validate our algorithm or their own algorithm. We believe that this approach can facilitate developing more accurate APCO algorithms and help in its clinical application. In the domain of detecting arrhythmia in the electrocardiogram, numerous studies and technologies have been proposed using publicly available data, such as the MIT-BIH dataset [
- ]. Likewise, our open dataset can be an academic reference for the APCO domain. Finally, a transfer learning method was proposed based on 2 datasets with different characteristics in bio-signal fields, and its scalability was confirmed. These techniques will provide good technical strategies for developing machine learning algorithms in the medical field with scanty data, such as PAC-based CO.Comparison With Prior Work
In a previous study, Moon et al [
] built a deep learning–based APCO algorithm using the data of 31 liver transplantation patients. However, their model only included the patients who underwent liver transplantation and has not been validated in the other types of surgery. In this study, a larger dataset was used that included both cardiac surgery and liver transplantation cases, with balanced ratios. In addition, a transfer learning technique was used, in which the parameters were pretrained with a large amount of APCO data and then tuned with PAC data. This may explain why our model worked better than the FloTrac for both cardiac and liver transplantation cases.Limitations
This study has some limitations. First, the data used in this study were from a single-center registry of a surgical cohort, which may contain a limited range of CO. This problem can be overcome by adding more data. However, the clinical use of PAC is gradually decreasing; other modalities such as a Doppler flowmeter or echocardiogram are required. Second, continuous CO measurement methods used as ground-truth values in this study can be less accurate in certain situations, such as in rapid fluid administration, compared with the gold-standard intermittent thermodilution technique [
]. In addition, the delay time for processing in the Vigilance II monitor was not fully revealed [ ]. Third, there was a statistical difference in age between the training and testing sets. This was an inevitable problem that occurred because of the use of real-world clinical data based on a prospective registry. However, elderly patients may have isolated systolic hypertension, which may alter the arterial pressure waveforms and affect the results. Fourth, there was no visualization with explainable artificial intelligence algorithms of how the proposed algorithm produces the results [ , ]. A proposal for a method that can display an indication for high or low CO and SVR from a waveform would have great clinical benefit. Finally, although various technological methods have been adopted, our developed model may be a local optimum and not a global optimum. Therefore, the raw data of this study was disclosed, allowing other researchers to improve the model.Conclusions
In conclusion, an uncalibrated APCO algorithm was developed and validated using a CNN and a transfer learning technique. The performance of our model was better than that of current commercial, uncalibrated APCO devices. Further improvement of the open-source algorithm developed in this study may be helpful for estimating cardiac output accurately in clinical practice and optimizing high-risk patient care.
Acknowledgments
This work was supported by the Seoul National University Hospital (SNUH) research fund (grant 06-2006-2449); the National Research Foundation of Korea (NRF) grant funded by the Ministry of Science and ICT (MSIT), Korea (NRF-2018R1A5A1060031); and Ministry of Science and ICT (MSIT), Korea, under the Information Technology Research Center (ITRC) support program (IITP-2018-0-01833), supervised by the Institute for Information & Communications Technology Promotion (IITP).
Conflicts of Interest
None declared.
Details of delaying the pulmonary artery catheter cardiac output data.
DOCX File , 504 KB
The individual plots of the time course of the measured and predicted stroke volume in all 99 testing cases.
PDF File (Adobe PDF File), 13581 KBReferences
- Nepogodiev D, Martin J, Biccard B, Makupe A, Bhangu A, National Institute for Health Research Global Health Research Unit on Global Surgery. Global burden of postoperative death. Lancet 2019 Feb 02;393(10170):401. [CrossRef] [Medline]
- Giglio MT, Marucci M, Testini M, Brienza N. Goal-directed haemodynamic therapy and gastrointestinal complications in major surgery: a meta-analysis of randomized controlled trials. Br J Anaesth 2009 Nov;103(5):637-646 [FREE Full text] [CrossRef] [Medline]
- Stevenson LW, Tillisch JH. Maintenance of cardiac output with normal filling pressures in patients with dilated heart failure. Circulation 1986 Dec;74(6):1303-1308. [CrossRef] [Medline]
- Kern JW, Shoemaker WC. Meta-analysis of hemodynamic optimization in high-risk patients. Crit Care Med 2002 Aug;30(8):1686-1692. [CrossRef] [Medline]
- Drummond KE, Murphy E. Minimally invasive cardiac output monitors. Continuing Education in Anaesthesia Critical Care & Pain 2012 Feb;12(1):5-10. [CrossRef]
- Marik PE. Obituary: pulmonary artery catheter 1970 to 2013. Ann Intensive Care 2013;3(1):38. [CrossRef]
- Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 1996 Sep 18;276(11):889-897. [CrossRef] [Medline]
- Hattori K, Maeda T, Masubuchi T, Yoshikawa A, Ebuchi K, Morishima K, et al. Accuracy and Trending Ability of the Fourth-Generation FloTrac/Vigileo System in Patients With Low Cardiac Index. Journal of Cardiothoracic and Vascular Anesthesia 2017 Feb;31(1):99-104. [CrossRef]
- Sakka SG, Kozieras J, Thuemer O, van Hout N. Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 2007 Sep;99(3):337-342 [FREE Full text] [CrossRef] [Medline]
- Krizhevsky A, Sutskever I, Hinton GE. ImageNet classification with deep convolutional neural networks. Commun. ACM 2017 May 24;60(6):84-90. [CrossRef]
- Hannun AY, Rajpurkar P, Haghpanahi M, Tison GH, Bourn C, Turakhia MP, et al. Cardiologist-level arrhythmia detection and classification in ambulatory electrocardiograms using a deep neural network. Nat Med 2019 Jan 7;25(1):65-69. [CrossRef]
- Johnson AEW, Pollard TJ, Shen L, Lehman LH, Feng M, Ghassemi M, et al. MIMIC-III, a freely accessible critical care database. Sci Data 2016 May 24;3:160035 [FREE Full text] [CrossRef] [Medline]
- Lee H, Jung C. Vital Recorder-a free research tool for automatic recording of high-resolution time-synchronised physiological data from multiple anaesthesia devices. Sci Rep 2018 Dec 24;8(1):1527 [FREE Full text] [CrossRef] [Medline]
- Beaulieu-Jones B, Finlayson SG, Chivers C, Chen I, McDermott M, Kandola J, et al. Trends and Focus of Machine Learning Applications for Health Research. JAMA Netw Open 2019 Oct 25;2(10):e1914051. [CrossRef]
- Jiang F, Jiang Y, Zhi H, Dong Y, Li H, Ma S, et al. Artificial intelligence in healthcare: past, present and future. Stroke Vasc Neurol 2017 Dec;2(4):230-243 [FREE Full text] [CrossRef] [Medline]
- Faust O, Hagiwara Y, Hong TJ, Lih OS, Acharya UR. Deep learning for healthcare applications based on physiological signals: A review. Computer Methods and Programs in Biomedicine 2018 Jul;161:1-13. [CrossRef]
- Van Steenkiste G, van Loon G, Crevecoeur G. Transfer Learning in ECG Classification from Human to Horse Using a Novel Parallel Neural Network Architecture. Sci Rep 2020 Jan 13;10(1):186. [CrossRef]
- Murugesan B, Ravichandran V, Ram KS. ECGNet: Deep Network for Arrhythmia Classification. In: P. P, Joseph J, Shankaranarayana SM, Sivaprakasam M. ECGNet: Deep Network for Arrhythmia Classification. 2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Internet Rome: IEEE; 2018 Jun 11 Presented at: 2018 IEEE International Symposium on Medical Measurements and Applications (MeMeA); 11-13 June 2018; Rome, Italy. [CrossRef]
- Ju C, Gao D, Mane R, Tan B, Liu Y, Guan C. Federated Transfer Learning for EEG Signal Classification. In: Federated Transfer Learning for EEG Signal Classification. 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC) Internet Montreal, QC. Canada: IEEE; 2020 Presented at: 2020 42nd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC); 20-24 July 2020; 2020: Montreal, QC, Canada. [CrossRef]
- Zheng W, Lu B. Personalizing EEG-based affective models with transfer learning. 2016 Presented at: Proceedings of the twenty-fifth international joint conference on artificial intelligence; 9-15 July 2016; New York, NY, USA.
- Vital DB. URL: https://vitaldb.net [accessed 2021-06-07]
- Cleveland WS, Devlin SJ. Locally Weighted Regression: An Approach to Regression Analysis by Local Fitting. Journal of the American Statistical Association 1988 Sep;83(403):596-610. [CrossRef]
- Lazor MA, Pierce ET, Stanley GD, Cass JL, Halpern EF, Bode RH. Evaluation of the accuracy and response time of stat-mode continuous cardiac output. Journal of Cardiothoracic and Vascular Anesthesia 1997 Jun;11(4):432-436. [CrossRef]
- Lakhal K, Ehrmann S, Boulain T. Predictive performance of passive leg raising in patients with atrial fibrillation. Br J Anaesth 2016 Sep;117(3):399 [FREE Full text] [CrossRef] [Medline]
- Aboy M, McNames J, Thong T, Tsunami D, Ellenby M, Goldstein B. An Automatic Beat Detection Algorithm for Pressure Signals. IEEE Trans. Biomed. Eng 2005 Oct;52(10):1662-1670. [CrossRef]
- Szegedy C, Liu W, Jia Y, Sermanet P, Reed S, Anguelov D, et al. Going deeper with convolutions. 2015 Presented at: Proceedings of the IEEE conference on computer vision and pattern recognition; 2015; Boston, MA, USA. [CrossRef]
- Wang X, Girshick R, Gupta A, He K. Non-local neural networks. 2018 Presented at: Proceedings of the IEEE conference on computer vision and pattern recognition (CVPR); 2018; Salt Lake City, UT, USA p. 7794. [CrossRef]
- Ioffe S, Szegedy C. Batch Normalization: Accelerating Deep Network Training by Reducing Internal Covariate Shift. 2015 Presented at: Proceedings of the 32nd International Conference on Machine Learning, ICML 2015; 6-11 July 2015; Lille, France p. 448.
- Srivastava N, Hinton G, Krizhevsky A, Sutskever I, Salakhutdinov R. Dropout: a simple way to prevent neural networks from overfitting. J. Mach. Learn. Res 2014:1929-1958.
- Glorot X, Bengio Y. Understanding the difficulty of training deep feedforward neural networks. 2010 Presented at: Proceedings of the thirteenth international conference on artificial intelligence and statistics (AISTATS); 2010; Chia Laguna Resort, Sardinia, Italy, p. 249.
- Liu L, Jiang H, He P, Chen W, Liu X, Gao J, et al. On the Variance of the Adaptive Learning Rate and Beyond. : OpenReview.net; 2020 Presented at: International Conference on Learning Representations (ICLR); April 26-30, 2020; Addis Ababa, Ethiopia URL: https://openreview.net/forum?id=rkgz2aEKDr
- Zhang M, Lucas J, Ba J, Hinton G. Lookahead Optimizer: k steps forward, 1 step back. 2019 Presented at: Advances in Neural Information Processing Systems 32: Annual Conference on Neural Information Processing Systems 2019, NeurIPS 2019; December 8-14, 2019; Vancouver, BC, Canada URL: https://proceedings.neurips.cc/paper/2019/file/90fd4f88f588ae64038134f1eeaa023f-Paper.pdf
- APCONet: AI model for arterial pressure waveform derived cardiac output. GitHub. 2020 Sep 28. URL: https://github.com/hyunlimy/APCONet [accessed 2021-07-27]
- Bland JM, Altman DG. Agreement Between Methods of Measurement with Multiple Observations Per Individual. Journal of Biopharmaceutical Statistics 2007 Jul 05;17(4):571-582. [CrossRef]
- Saugel B, Grothe O, Wagner JY. Tracking Changes in Cardiac Output: Statistical Considerations on the 4-Quadrant Plot and the Polar Plot Methodology. Anesth Analg 2015 Aug;121(2):514-524. [CrossRef] [Medline]
- Nordström J, Hällsjö-Sander C, Shore R, Björne H. Stroke volume optimization in elective bowel surgery: a comparison between pulse power wave analysis (LiDCOrapid) and oesophageal Doppler (CardioQ). British Journal of Anaesthesia 2013 Mar;110(3):374-380. [CrossRef]
- Frank O. The basic shape of the arterial pulse. First treatise: mathematical analysis. 1899. J Mol Cell Cardiol 1990 Mar;22(3):255-277. [CrossRef] [Medline]
- Thiele RH, Durieux ME. Arterial Waveform Analysis for the Anesthesiologist. Anesthesia & Analgesia 2011;113(4):766-776. [CrossRef]
- Sun JX, Reisner AT, Saeed M, Heldt T, Mark RG. The cardiac output from blood pressure algorithms trial. Critical Care Medicine 2009;37(1):72-80. [CrossRef]
- Rhodes A, Sunderland R. Arterial Pulse Power Analysis: The LiDCOℳ plus System. In: Update in Intensive Care and Emergency Medicine. Berlin, Heidelberg: Springer; 2005:183.
- Gödje O, Höke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, et al. Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 2002 Jan;30(1):52-58. [CrossRef] [Medline]
- Hofer CK, Senn A, Weibel L, Zollinger A. Assessment of stroke volume variation for prediction of fluid responsiveness using the modified FloTrac and PiCCOplus system. Crit Care 2008;12(3):R82 [FREE Full text] [CrossRef] [Medline]
- Alhashemi JA, Cecconi M, Hofer CK. Cardiac output monitoring: an integrative perspective. Crit Care 2011;15(2):214. [CrossRef]
- Moody G, Mark R. The impact of the MIT-BIH Arrhythmia Database. IEEE Eng. Med. Biol. Mag 2001;20(3):45-50. [CrossRef]
- Mousavi S, Afghah F. Inter- and Intra- Patient {ECG} Heartbeat Classification for Arrhythmia Detection: {A} Sequence to Sequence Deep Learning Approach. ICASSP 2019 - 2019 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP) Internet Brighton, United Kingdom: IEEE; 2019 Presented at: IEEE International Conference on Acoustics, Speech and Signal Processing, ICASSP; May 12-17, 2019; Brighton, United Kingdom. [CrossRef]
- Li R, Zhang X, Dai H, Zhou B, Wang Z. Interpretability Analysis of Heartbeat Classification Based on Heartbeat Activity’s Global Sequence Features and BiLSTM-Attention Neural Network. IEEE Access 2019;7:109870-109883. [CrossRef]
- Kachuee M, Fazeli S, Sarrafzadeh M. ECG Heartbeat Classification: A Deep Transferable Representation. 2018 Presented at: IEEE International Conference on Healthcare Informatics, ICHI; June 4-7, 2018; New York City, NY, USA p. A. [CrossRef]
- Moon Y, Moon HS, Kim D, Kim J, Lee J, Shim W, et al. Deep Learning-Based Stroke Volume Estimation Outperforms Conventional Arterial Contour Method in Patients with Hemodynamic Instability. J Clin Med 2019 Sep 09;8(9):1419 [FREE Full text] [CrossRef] [Medline]
- Slagt C, Malagon I, Groeneveld ABJ. Systematic review of uncalibrated arterial pressure waveform analysis to determine cardiac output and stroke volume variation. Br J Anaesth 2014 Apr;112(4):626-637 [FREE Full text] [CrossRef] [Medline]
- Selvaraju RR, Cogswell M, Das A, Vedantam R, Parikh D, Batra D. Grad-CAM: Visual Explanations from Deep Networks via Gradient-Based Localization. Int J Comput Vis 2019 Oct 11;128(2):336-359. [CrossRef]
- Ribeiro M, Singh S, Guestrin C. "Why Should I Trust You?": Explaining the Predictions of Any Classifier. : ACM; 2016 Presented at: Proceedings of the 22nd ACM SIGKDD International Conference on Knowledge Discovery and Data Mining; August 13-17, 2016; San Francisco, CA, USA p. 1135. [CrossRef]
Abbreviations
APCO: arterial pressure-based cardiac output |
CNN: convolutional neural network |
CO: cardiac output |
HR: heart rate |
LOWESS: locally weighted scatterplot smoother |
PAC: pulmonary artery catheter |
SV: stroke volume |
SVR: systemic vascular resistance |
Edited by C Lovis; submitted 04.10.20; peer-reviewed by D Pförringer, J Kim, L Rusu; comments to author 17.12.20; revised version received 10.02.21; accepted 17.06.21; published 16.08.21
Copyright©Hyun-Lim Yang, Chul-Woo Jung, Seong Mi Yang, Min-Soo Kim, Sungho Shim, Kook Hyun Lee, Hyung-Chul Lee. Originally published in JMIR Medical Informatics (https://medinform.jmir.org), 16.08.2021.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Medical Informatics, is properly cited. The complete bibliographic information, a link to the original publication on https://medinform.jmir.org/, as well as this copyright and license information must be included.