Original Paper
Abstract
Background: Critically ill patients require constant point-of-care blood glucose testing to guide insulin-related decisions. Transcribing these values from glucometers into a paper log and the electronic medical record is very common yet error-prone in intensive care units, given the lack of connectivity between glucometers and the electronic medical record in many US hospitals.
Objective: We examined (1) transcription errors of glucometer blood glucose values documented in the paper log and in the electronic medical record vital signs flow sheet in a surgical trauma intensive care unit, (2) insulin errors resulting from transcription errors, (3) lack of documenting these values in the paper log and the electronic medical record vital signs flow sheet, and (4) average time for docking the glucometer.
Methods: This secondary data analysis examined 5049 point-of-care blood glucose tests. We obtained values of blood glucose tests from bidirectional interface software that transfers the meters’ data to the electronic medical record, the paper log, and the vital signs flow sheet. We obtained patient demographic and clinical-related information from the electronic medical record.
Results: Of the 5049 blood glucose tests, which were pertinent to 234 patients, the total numbers of undocumented or untranscribed tests were 608 (12.04%) in the paper log, 2064 (40.88%) in the flow sheet, and 239 (4.73%) in both. The numbers of transcription errors for the documented tests were 98 (2.21% of 4441 documented tests) in the paper log, 242 (8.11% of 2985 tests) in the flow sheet, and 43 (1.64% of 2616 tests) in both. The numbers of transcription errors per patient were 0.4 (98 errors/234 patients) in the paper log, 1 (242 errors/234 patients) in the flow sheet, and 0.2 in both (43 errors/234 patients). Transcription errors in the paper log, the flow sheet, and in both resulted in 8, 24, and 2 insulin errors, respectively. As a consequence, patients were given a lower or higher insulin dose than the dose they should have received had there been no errors. Discrepancies in insulin doses were 2 to 8 U lower doses in paper log transcription errors, 10 U lower to 3 U higher doses in flow sheet transcription errors, and 2 U lower in transcription errors in both. Overall, 30 unique insulin errors affected 25 of 234 patients (10.7%). The average time from point-of-care testing to meter docking was 8 hours (median 5.5 hours), with some taking 56 hours (2.3 days) to be uploaded.
Conclusions: Given the high dependence on glucometers for point-of-care blood glucose testing in intensive care units, full electronic medical record-glucometer interoperability is required for complete, accurate, and timely documentation of blood glucose values and elimination of transcription errors and the subsequent insulin-related errors in intensive care units.
doi:10.2196/11873
Keywords
Introduction
Background
Glycemic control in critically ill patients is essential to improve clinical outcomes and decrease morbidity and mortality [
- ], specifically for patients admitted to intensive care units (ICUs) for more than 3 days [ ] and for patients admitted to surgical trauma ICUs (STICUs) compared with medical ICUs [ ]. Critically ill patients require constant point-of-care tests (POCTs) for blood glucose to guide initiation and titration decisions regarding continuous insulin infusion following insulin management protocols. Handheld blood glucose monitoring devices or glucometers are widely used in ICUs for this purpose for convenience and portability [ , ].Transcribing blood glucose readings from glucometers into a paper log and different flow sheets in the electronic medical record (EMR) by health care professionals is a very common yet error-prone practice in ICUs, given the lack of interoperability or connectivity between glucometers and the EMR in many US hospitals [
]. Interoperability allows for wireless transfer of blood glucose values from glucometers to the EMR without the need for manual data entry. Despite the call for system interoperability and emerging research describing frameworks and prototypes for seamless integration of medical device data into the EMR using different connectivity standards [ - ], medical device-EMR connectivity is limited in the United States. In a national survey of 825 US hospitals, the Health Information and Management Systems Society Analytics team reported a lack of any interface between EMRs and medical devices in 70% of the hospitals. The remaining 30% of the hospitals reported an interface of an average of 2.6 device types (out of 11 devices) with their EMRs. Interestingly, none of the hospitals provided an interface between glucometers and the EMR [ ].Extensive literature exists on the use of glucometers in ICUs. However, most studies focused on the glucometers’ accuracy in comparison with other blood glucose analytical measures [
- ]. Research on transcription errors is also available [ - ]; however, there is a paucity of research on transcription errors of blood glucose values obtained by glucometers into the EMRs and the subsequent insulin errors [ ]. Although the use of glucometers with high specificity and sensitivity is essential in critical care settings to prevent harmful effects of erroneous blood glucose readings and the subsequent underdose or overdose of insulin therapy, accurate and instant documentation of blood glucose values obtained by glucometers into the EMR is equally important to inform glycemic control and insulin management decisions.Objective
This study examined (1) transcription errors of blood glucose values obtained by a glucometer that were documented in the paper log by technicians and in the EMR vital signs flow sheet by nurses in the ICU, (2) insulin errors resulting from transcription errors of blood glucose values, (3) lack of documenting blood glucose values in the paper log and the EMR vital signs flow sheet, and (4) meter docking time.
Methods
Design, Sample, and Setting
This secondary data analysis study examined 5049 blood glucose tests for transcription errors, insulin errors, lack of documenting blood glucose values in the paper log and the EMR, and meter docking time. The study took place in a 30-bed STICU located in a 705-bed university teaching hospital with a large referral base in the southwestern United States. The STICU has an annual admission rate of 1600 patients and an approximate monthly admission rate of 133 patients. At the time of the study, there were 46 full-time and 11 part-time nurses and 13 technicians working in the unit. The average range of blood glucose POCTs performed on patients in the unit is 4200 to 4300 tests per month.
After obtaining institutional review board approval from the University of Texas and the University Health System (number 20140330H), we performed the audit of blood glucose tests and insulin data in a 20% stratified sample of all blood glucose tests available in the meters for patients admitted during 4 months (July to October 2016). Stratification was based on the working shift (day or night) as the only possible factor that may introduce transcription errors of blood glucose readings as a result of fatigue expected at the end of each working shift and on the night shift. Additionally, when we selected a blood glucose test, we also included all blood glucose tests pertinent to the same patient within the same episode of admission to evaluate errors per patient. This resulted in a total of 5049 blood glucose tests.
Description of the Point-of-Care Testing of Blood Glucose
The point-of-care glucose testing device is Accu-Chek Inform II (Roche Diagnostics Corporation, Indianapolis, IN, USA).
depicts a functional workflow model for this process of testing. The process starts by the physician ordering a POCT. The nurse informs the technician about the order, who in turn performs the test using the glucometer and transcribes the result into a paper log—a grid that includes the patient’s name, visit identification number (VIN), room number, time and date of the test, and the result. The VIN is a unique number for each patient episode of admission that is obtained by scanning the patient’s wristband at the time of performing the test.Nurses then manually enter the readings for each patient into the EMR vital signs flow sheet and use this information to inform their insulin management decisions following physician orders and insulin management protocols. Clinical decisions include whether to continue to monitor, repeat the test to verify critical blood glucose values, inform the physician, give insulin, and titrate the insulin drip based on the insulin management protocol. The blood glucose values entered by nurses into the EMR vital signs flow sheet can be obtained (1) from the glucometer itself by manually searching the readings using the time of the test and the patient VIN to locate the test value, (2) from the technician, who verbally endorses the value to the nurse if he or she is available in the unit, or (3) by checking the value transcribed by the technician into the paper log.
The technician docks the meter by placing it into the meter base unit within 24 hours after the time of the first test for a given day. Meters maintain log data for up to 2000 readings. Since the meter can be docked after 24 hours of use, nurses usually base their insulin management decisions on the readings transcribed by the technicians into the paper log or the readings entered by the nurses into the vital signs flow sheet. By docking the meter, readings are automatically uploaded into the RALS-Plus database, which interfaces with the EMR laboratory flow sheet. These data include the examiner’s employee identification number, patient identification (name, VIN), date and time of the test, time the meter was docked, and blood glucose values. It is worth noting that there is no direct link or seamless transfer of data in the EMR between the vital signs flow sheet and the EMR laboratory flow sheet.
The RALS-Plus v1.5.1 (Alere North America, LLC, Orlando, FL, USA) is a bidirectional interface software for in-hospital glucometers that uploads meter data into the EMR laboratory flow sheet only. The software also generates different types of reports for quality improvement. Data can be generated based on criteria such as the start and end date of the test, blood glucose values, patient VIN, sample type, and test location. Reports can be emailed, printed, saved, or exported into an Excel, rich text (rtf), or pdf file format.
Main Outcome Variables
Transcription Errors
Since the focus of this study was transcription errors, we assumed that technicians follow best practices in obtaining blood samples and in meter use according to the unit policies and procedures and the glucometer’s user manual. Blood glucose values uploaded into RALS-Plus are those in the meters and they are transferred to the EMR laboratory flow sheet. These values are accurate. Transcribing blood glucose values from the meters to the paper log and the EMR vital signs flow sheet may result in 3 potential types of errors (
). The “corresponding values” ( ) in the paper log and the EMR vital signs flow sheet are based on the same patient VIN, same date of the test, and within a 1-hour time frame from the POCT (time in RALS) to the time the test was transcribed into the paper log or the EMR vital signs flow sheet.Paper log | EMR vital signs flow sheet | |
Flow sheet correct | Flow sheet wrong | |
Paper correct | No error: The blood glucose value in the RALS database matches the corresponding value transcribed by technicians and nurses into the paper log and the EMR vital signs flow sheet for a given test in a given date and time. | Vital signs flow sheet error: Any discrepancy regardless of the magnitude between blood glucose value in the RALS database and the corresponding value transcribed by nurses into the EMR vital signs flow sheet. |
Paper wrong | Paper log error: Any discrepancy regardless of the magnitude between blood glucose value in the RALS database and the corresponding value transcribed by the technician into the paper log. | Paper log and vital signs flow sheet error: The 2 blood glucose values transcribed by technicians and nurses into the paper log and the EMR vital signs flow sheet for a given test in a given date and time do not match the value in the RALS database. |
Undocumented Values of Blood Glucose Tests
Untranscribed or undocumented blood glucose values are those available in RALS database but were not transcribed into the paper log or entered into the EMR vital signs flow sheet.
Insulin Errors Related to Erroneously Transcribed Blood Glucose Values
For each transcription error, we also examined whether that error resulted in giving the wrong dose of insulin. We evaluated the wrong insulin dose based on administering a higher or lower insulin dose, regardless of the magnitude of the difference, than the one recommended by the protocol for the correct blood glucose value (the value in the RALS system) or not giving insulin when it should be administered to the patient according to the insulin management protocol based on the correct blood glucose value.
Meter Docking Time
As mentioned above, we considered a 1-hour time frame from the POCT (time in RALS or glucometer) to the time the test result was transcribed into the paper log or the EMR vital signs flow sheet when we retrieved the time for transcribing blood glucose values. Meter docking time was retrieved from the RALS database and is the time from the POCT to the time meters were docked (readings were uploaded into the EMR laboratory flow sheet).
In addition to these outcomes, we also collected patient demographics and clinical-related information such as age, sex, diagnosis, diabetes status, admission and discharge dates, and total number of POCTs the patient underwent during the ICU stay.
Data Collection Procedure
We took the following steps in the sequence identified to collect the data. Three nurse educators collected the data from the paper log and the EMR vital signs flow sheet to enhance objectivity.
First, we accessed the RALS database for the selected study months and downloaded the Excel file (Microsoft Corporation). The file included the patient’s name, VIN, EMR number, test date and time, blood glucose value (meter value), and time of docking the meter.
Second, we selected a stratified sample of 20% of the blood glucose readings and the related information from RALS from the Excel file. In addition to the 20% sample of readings, we went back and selected all pertinent blood glucose tests within the episode of admission for all VINs included in the stratified sample.
Third, for each test selected from RALS, we accessed the EMR and obtained patient demographics and clinical-related information based on the VIN, as well as the corresponding values of blood glucose transcribed into the vital signs flow sheet and time of documentation. We also accessed the laboratory flow sheet to make sure that the tests in RALS were pertinent to that patient.
Fourth, for each test selected from RALS (step 2) for each patient and based on the VIN, we accessed the paper log using the patient’s name and VIN as the identifiers. We obtained the corresponding blood glucose value for each test using the date and a 1-hour time frame from the POCT (time in RALS) as the matching codes. We also obtained the actual time of the test documented in the paper log.
Data Analysis
We used R statistical computing software v3.5.1 (R Foundation) to analyze the data. Patients’ characteristics and all types of errors were presented using descriptive statistics. We examined the difference in number of POCTs between diabetic and nondiabetic patients using Student t test with significance set at P<.05.
We limited the analysis of transcription errors to cases where the results of the blood glucose tests were transcribed by clinicians and nurses. For example, the denominator for the paper log transcription errors was the number of blood glucose readings transcribed into the paper log, excluding missing values (ie, when the readings were not transcribed).
Results
Patient Characteristics and Number of Point-of-Care Tests
The 5049 blood glucose tests analyzed for transcription errors, undocumented blood glucose readings, and meter docking time were pertinent to 234 unique patients, each with a unique VIN.
presents the patients’ characteristics. Most of the patients with documented diabetes status in the dataset did not have diabetes (93/234). Of the 234 patients, 97 were with unknown diabetes status. The average number of POCTs performed on diabetic patients ( ) was significantly higher than on nondiabetic patients (t47=–2.17, P=.03). One of the patients had 792 POCTs during his stay ( ). The median number of POCTs for diabetic patients was 12 tests.Patient characteristic | Statistics | |
Age in years, mean (SD) | 57.5 (17.4) | |
Length of stay in days, mean (SD) | 24.8 (48.3) | |
Number of point-of-care tests per patient, mean (SD) | 25.5 (67.9) | |
Male sex, n (%) | 131 (56.0) | |
Diabetes status, n (%) | ||
Yes | 44 (32.1) | |
No | 93 (67.9) | |
Missing | 97 (41.5) |
Diabetes status | Minimum | Median | Mean (SD) | Maximum |
Diabetes (n=44) | 1 | 12 | 60 (126) | 792 |
No diabetes (n=93) | 1 | 6 | 19 (40) | 344 |
Missing Documentation and Transcription Errors
describes the number of tests that were not transcribed into the paper log or the EMR vital signs flow sheet, or both, as well as the number of transcription errors. In the vital signs flow sheet, 40.88% of the tests (2064/5049 tests) were not transcribed. Of the blood glucose tests, 4.73% (239/5049 tests) were not transcribed in the paper log and in the EMR vital signs flow sheet at the same time.
Source | Undocumented tests, n (%) | Tests analyzed for errors, n | Errors among tests analyzed, n (%) | Range of error, mg/dL (mmol/L)a | Error rate per patient (N=234) |
Paper log | 608 (12.04) | 4441 | 98 (2.21) | –92 to 92 (–5.1 to 5.1) | 0.4 (98/234) |
Vital signs flow sheet | 2064 (40.88) | 2985 | 242 (8.11) | –110 to 80 (–6.1 to 4.4) | 1.0 (242/234) |
Both | 239 (4.73) | 2616 | 43 (1.64) | N/Ab | 0.2 (43/234) |
aRange of difference between the correct blood glucose value and the erroneously transcribed value.
bN/A: not applicable.
We analyzed all types of transcription errors when the blood glucose value was transcribed (
). Of the transcription errors among the 4441 transcribed tests in the paper logs, there were 98 (2.21%) errors. These errors were related to 30 of the 234 patients (12.8%). Of the 2985 transcribed values in the vital signs flow sheet, there were 242 (8.11%) errors related to 63 of the 234 patients (26.9%). The total number of paper log and vital signs flow sheet transcription errors among the 2616 tests analyzed was 43 (1.64%), related to 24 of the 234 patients (10.3%). Overall, among the 234 patients, there were 68 (29.1%) unique patients involved in all types of errors.Errors in the paper log resulted in transcribing a blood glucose value that was up to 92 mg/dL (5.1 mmol/L) lower or 92 mg/dL (5.1 mmol/L) higher than the correct value (the one in the EMR laboratory flow sheet or RALS). However, most errors, those between the 25th and 75th percentiles, were 12 mg/dL (0.7 mmol/L) lower to 7 mg/dL (0.4 mmol/L) higher than the accurate value. In the EMR vital signs flow sheet, the difference between the correct blood glucose value and the erroneously transcribed value was 110 mg/dL (6.1 mmol/L) lower to 80 mg/dL (4.4 mmol/L) higher. Most errors, those between the 25th and 75th percentiles, were 3 mg/dL (0.16 mmol/L) lower to 4 mg/dL (0.2 mmol/L) higher than the accurate values.
There were no significant differences in the number of transcription errors between the day shift and night shift (
).Source | Total errors, n (%) | Day shift, n (%) | Night shift, n (%) | Chi-square | df | P value |
Paper log | 98/4441 (2.21) | 53/2790 (1.90) | 45/1651 (2.73) | 2.9 | 1 | .09 |
Vital signs flow sheet | 242/2985 (8.11) | 163/1847 (8.83) | 79/1138 (6.94) | 3.1 | 1 | .08 |
Both | 43/2616 (1.64) | 24/1639 (1.46) | 19/977 (1.94) | 0.6 | 1 | .44 |
Insulin Errors
The 242 transcription errors in the EMR vital signs flow sheet resulted in 24 insulin errors. These errors resulted in giving 10 U lower to 3 U higher insulin dose than the dose that should have been given had there been no transcription errors. The 98 transcription errors in the paper log resulted in 8 insulin errors and giving 2 to 8 U lower insulin dose than the dose that should have been given had there been no transcription errors. The 43 errors in the EMR vital signs flow sheet and paper logs resulted in 2 insulin errors, both with 2 U lower than the correct insulin dose. Overall, there were 30 unique insulin errors that affected 25 of the 234 patients (10.7%).
Documentation Time
The average time from the POCT to the time meters were docked (readings were uploaded into the EMR laboratory flow sheet) was 8 hours with a median of 5.5 hours. Most readings, between the first and the third quartiles, took 1.3 to 12 hours to be uploaded into the EMR laboratory flow sheet. Some of the readings took 56 hours (2.3 days) to be uploaded into the EMR laboratory flow sheet.
In addition to these outcomes, we found 40 readings that were documented to some patients’ EMRs and the paper log after the date of discharge.
Discussion
Principal Findings
This study examined transcription errors of blood glucose tests obtained by a glucometer and documented in the paper log by ICU technicians and in the EMR vital signs flow sheet by ICU nurses. Insulin errors resulted from transcription errors of blood glucose values, the number of undocumented blood glucose tests in the paper log and the EMR vital signs flow sheet, and the average meter docking time. Research on the use of glucometers in ICU and non-ICU settings is extensive. However, most of these studies focused on precision and accuracy of the glucometers, sources of glucometer measurement errors, and the difference in sensitivity and specificity between glucometer devices from different vendors [
- ]. Nevertheless, glucometers are commonly used handheld devices to measure blood glucose at the point of care, specifically in ICUs to inform timely clinical decisions regarding insulin therapy. To our knowledge, this is the first study to examine transcription errors of blood glucose tests obtained by glucometers and to focus on the urgent need for EMR-glucometer interoperability.Transcription errors ranged from 2% for paper log errors to 8% for vital signs flow sheet errors. These errors resulted in a total of 30 insulin errors and affected 11% of the patients. The higher percentage of transcription errors in the vital signs flow sheet than in the paper log might be explained by a clinical workflow that has nurses obtain the results of blood glucose tests from 3 different sources, which are the paper log, the technicians, or the glucometers, while the technicians obtain the values only from the glucometers. Transcription errors in the vital signs flow sheet are clinically more significant than transcription errors in the paper log because they inform nurses’ insulin management decisions. These errors affected 63 (27%) patients.
It is important to note that we examined transcription errors and the associated insulin errors only when the blood glucose test results were transcribed by technicians and nurses. The very high percentage of untranscribed values (ie, up to 41% untranscribed into the vital signs flow sheet, n=2064) could mask the actual rate of transcription errors. Possible explanations for not transcribing blood glucose values might be workload issues and the assumption that all readings eventually will be available in the laboratory flow sheet in the EMR after docking the meter. In addition, finding 40 readings documented to some patients’ EMRs and the paper log after the date of discharge is alarming. This means that technicians were not scanning the patient bracelet but probably a sticker that remained on the paper log or the patient monitor or bed. Although eliminating the use of a paper log via a full EMR-glucometer interoperability could decrease this error, adherence to the unit policies and procedures for safe testing is critical for complete elimination of this error.
Although a partial interface exists in our hospital between glucometers and the EMR through the RALS bidirectional interface software, this interface transfers the data only to the EMR central laboratory flow sheet. Additionally, based on the unit policies and procedures, the meters should be docked within 24 hours by technicians. This long time period hinders the availability of the tests’ values at the point of care, making these data unusable for immediate clinical decisions. Furthermore, our results showed that, in reality, docking the meters might take up to more than 2 days. Therefore, there is an urgent need for full glucometer-EMR connectivity to allow for seamless transfer of meter data into other fields of the EMR (ie, the vital signs flow sheet) in order to eliminate data transcription errors and the associated insulin errors.
The few available studies on medical devices-EMR connectivity have focused on vital signs monitors in ICUs and supported improved efficiency and elimination of transcription errors when vital signs monitoring devices are connected to the EMR [
]. The results of our study support the urgent need for a comprehensive and instant connectivity to transfer glucometer data to all fields of the EMR to better inform clinical decisions and eliminate insulin errors associated with transcription errors. On the other hand, from an engineering perspective, interoperability challenges do exist. These may include lack of research describing successes and challenges, the complexity of data elements, and the difference in type of information and formats in which information is stored and displayed. Most important, studies supported the potential for new types of errors in device connectivity, such as transferring the data into the wrong patient’s EMR, in addition to the slow speed of the interface attributed to the slow speed of older medical devices and computers [ ]. Therefore, the process of and errors associated with interoperability should be carefully examined.Limitations
The results of this study should be interpreted in light of the following limitations. First, since workload, admission rate, and the large number of monthly POCTs are inherent factors that may affect transcription errors, our results can only be generalized to STICUs with a similar workload and rate of POCTs. Second, we limited the errors examined in this study to transcription errors; measurement errors of blood glucose values that may result from inappropriate testing or scanning the wrong patients were beyond the scope of this study. Third, because we collected retrospective data, our risk assessment was limited to identifying the number of insulin errors resulting from transcription errors without identifying the clinical consequences or adverse events of insulin errors. On the other hand, insulin is a high-alert medication and errors in its administration may cause serious hypoglycemia and hyperglycemia, seizures, coma, ketoacidosis, and even death [
].Conclusions
Transcription errors of blood glucose values obtained by glucometers do exist and result in insulin errors. Given the high dependence on glucometers for POCTs of blood glucose in ICUs, full EMR-glucometer interoperability is required for complete and accurate documentation of blood glucose values, and elimination of transcription errors and the subsequent insulin-related errors in ICUs.
Acknowledgments
This project was funded by the University Health System’s Center for Clinical Excellence and the University of Texas Health School of Nursing. The role of the funding sources was limited to providing financial support for the conduct of the research. Sponsors were not involved in study design; collection, analysis, or interpretation of data; writing of the manuscript; or the decision to submit the article for publication.
Conflicts of Interest
None declared.
References
- Krinsley JS, Maurer P, Holewinski S, Hayes R, McComsey D, Umpierrez GE, et al. Glucose control, diabetes status, and mortality in critically ill patients: the continuum from intensive care unit admission to hospital discharge. Mayo Clin Proc 2017 Dec;92(7):1019-1029. [CrossRef] [Medline]
- Van den Berghe G, Wilmer A, Hermans G, Meersseman W, Wouters PJ, Milants I, et al. Intensive insulin therapy in the medical ICU. N Engl J Med 2006 Feb 02;354(5):449-461. [CrossRef] [Medline]
- Silva-Perez LJ, Benitez-Lopez MA, Varon J, Surani S. Management of critically ill patients with diabetes. World J Diabetes 2017 Mar 15;8(3):89-96 [FREE Full text] [CrossRef] [Medline]
- Williams JB, Peterson ED, Albrecht ÁS, Li S, Hirji SA, Ferguson T, et al. Glycemic control in patients undergoing coronary artery bypass graft surgery: clinical features, predictors, and outcomes. J Crit Care 2017 Dec;42:328-333. [CrossRef] [Medline]
- Murphy CV, Coffey R, Cook CH, Gerlach AT, Miller SF. Early glycemic control in critically ill patients with burn injury. J Burn Care Res 2011;32(6):583-590. [CrossRef] [Medline]
- Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc 2003 Dec;78(12):1471-1478. [CrossRef] [Medline]
- Griesdale DEG, de Souza RJ, van Dam RM, Heyland DK, Cook DJ, Malhotra A, et al. Intensive insulin therapy and mortality among critically ill patients: a meta-analysis including NICE-SUGAR study data. CMAJ 2009 Apr 14;180(8):821-827 [FREE Full text] [CrossRef] [Medline]
- Wiener RS, Wiener DC, Larson RJ. Benefits and risks of tight glucose control in critically ill adults: a meta-analysis. JAMA 2008 Aug 27;300(8):933-944. [CrossRef] [Medline]
- Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol 2009 Jul 01;3(4):971-980 [FREE Full text] [CrossRef] [Medline]
- Klonoff DC. Point-of-care blood glucose meter accuracy in the hospital setting. Diabetes Spectr 2014 Aug;27(3):174-179 [FREE Full text] [CrossRef] [Medline]
- HIMSS. Medical devices landscape: current and future adoption, integration with EMRs, and connectivity. A HIMSS Analytics white paper. Chicago, IL: HIMSS Analytics; 2010. URL: https://www.lantronix.com/wp-content/uploads/pdf/Medical-Devices-Landscape_Lantonix_HIMMS_WP.pdf [accessed 2018-08-06] [WebCite Cache]
- Choi JS, Lee JH, Park JH, Nam HS, Kwon H, Kim D, et al. Design and implementation of a seamless and comprehensive integrated medical device interface system for outpatient electronic medical records in a general hospital. Int J Med Inform 2011 Apr;80(4):274-285. [CrossRef] [Medline]
- Medical device plug-and-play (MD PnP) interoperability program. Cambridge, MA: MD PnP; 2012 Oct. URL: http://www.mdpnp.org/uploads/Oct12_MD_PnP_White_Paper.pdf [accessed 2018-08-06] [WebCite Cache]
- Fonseca T, Ribeiro C, Granja C. Vital signs in intensive care: automatic acquisition and consolidation into electronic patient records. J Med Syst 2009 Feb;33(1):47-57. [Medline]
- Mense A, Sauermann S, Gerbovic G, Frohner M, Pohn B, Bruckner R, et al. Usage of international standards for integrating extramural monitoring and personal health device data into medical information infrastructure. Stud Health Technol Inform 2010;160(Pt 1):550-553. [Medline]
- Mahmoodpoor A, Hamishehkar H, Shadvar K, Sanaie S, Iranpour A, Fattahi V. Validity of bedside blood glucose measurement in critically ill patients with intensive insulin therapy. Indian J Crit Care Med 2016 Nov;20(11):653-657 [FREE Full text] [CrossRef] [Medline]
- Zhang R, Isakow W, Kollef MH, Scott MG. Performance of a modern glucose meter in ICU and general hospital inpatients: 3 years of real-world paired meter and central laboratory results. Crit Care Med 2017 Sep;45(9):1509-1514. [CrossRef] [Medline]
- DuBois JA, Slingerland RJ, Fokkert M, Roman A, Tran NK, Clarke W, et al. Bedside glucose monitoring-is it safe? A new, regulatory-compliant risk assessment evaluation protocol in critically ill patient care settings. Crit Care Med 2017 Apr;45(4):567-574 [FREE Full text] [CrossRef] [Medline]
- Joseph JI. Analysis: new point-of-care blood glucose monitoring system for the hospital demonstrates satisfactory analytical accuracy using blood from critically ill patients--an important step toward improved blood glucose control in the hospital. J Diabetes Sci Technol 2013 Sep 01;7(5):1288-1293 [FREE Full text] [CrossRef] [Medline]
- Inoue S, Egi M, Kotani J, Morita K. Accuracy of blood-glucose measurements using glucose meters and arterial blood gas analyzers in critically ill adult patients: systematic review. Crit Care 2013 Mar 18;17(2):R48 [FREE Full text] [CrossRef] [Medline]
- Slater-MacLean L, Cembrowski G, Chin D, Shalapay C, Binette T, Hegadoren K, et al. Accuracy of glycemic measurements in the critically ill. Diabetes Technol Ther 2008 Jun;10(3):169-177. [CrossRef] [Medline]
- Mraovic B, Schwenk ES, Epstein RH. Intraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements. J Diabetes Sci Technol 2012 May 01;6(3):541-546 [FREE Full text] [CrossRef] [Medline]
- Hoedemaekers CWE, Klein Gunnewiek JMT, Prinsen MA, Willems JL, Van der Hoeven JG. Accuracy of bedside glucose measurement from three glucometers in critically ill patients. Crit Care Med 2008 Nov;36(11):3062-3066. [CrossRef] [Medline]
- American Association of Diabetic Educators. Practice advisory blood glucose meter accuracy. Chicago, IL: AADE; 2013 Sep 16. URL: https://www.diabeteseducator.org/docs/default-source/legacy-docs/_resources/pdf/research/Practice_Advisory_BGM_FINAL.pdf [accessed 2018-08-07] [WebCite Cache]
- MacKay M, Anderson C, Boehme S, Cash J, Zobell J. Frequency and severity of parenteral nutrition medication errors at a large children’s hospital after implementation of electronic ordering and compounding. Nutr Clin Pract 2016 Apr;31(2):195-206. [CrossRef] [Medline]
- South DA, Skelley JW, Dang M, Woolley T. Near-miss transcription errors: a comparison of reporting rates between a novel error-reporting mechanism and a current formal reporting system. Hosp Pharm 2015 Feb;50(2):118-124 [FREE Full text] [CrossRef] [Medline]
- Elsaid K, Truong T, Monckeberg M, McCarthy H, Butera J, Collins C. Impact of electronic chemotherapy order forms on prescribing errors at an urban medical center: results from an interrupted time-series analysis. Int J Qual Health Care 2013 Dec;25(6):656-663. [CrossRef] [Medline]
- Campion TR, May AK, Waitman LR, Ozdas A, Gadd CS. Effects of blood glucose transcription mismatches on a computer-based intensive insulin therapy protocol. Intensive Care Med 2010 Sep;36(9):1566-1570 [FREE Full text] [CrossRef] [Medline]
- Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009 Jul 01;3(4):903-913 [FREE Full text] [CrossRef] [Medline]
- Prescrire Editorial Staff. Insulin use: preventable errors. Prescrire Int 2014 Jan;23(145):14-17. [Medline]
Abbreviations
EMR: electronic medical record |
ICU: intensive care unit |
POCT: point-of-care test |
STICU: surgical trauma intensive care unit |
VIN: visit identification number |
Edited by C Lovis; submitted 08.08.18; peer-reviewed by K Glover, L Katz; comments to author 27.12.18; revised version received 10.01.19; accepted 11.01.19; published 25.03.19
Copyright©Azizeh Khaled Sowan, Ana Vera, Ashwin Malshe, Charles Reed. Originally published in JMIR Medical Informatics (http://medinform.jmir.org), 25.03.2019.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Medical Informatics, is properly cited. The complete bibliographic information, a link to the original publication on http://medinform.jmir.org/, as well as this copyright and license information must be included.