Maintenance Notice

Due to necessary scheduled maintenance, the JMIR Publications website will be unavailable from Wednesday, July 01, 2020 at 8:00 PM to 10:00 PM EST. We apologize in advance for any inconvenience this may cause you.

Who will be affected?

Advertisement

Previously submitted to: JMIR Medical Informatics (no longer under consideration since Jan 12, 2020)

Date Submitted: Nov 12, 2019
Open Peer Review Period: Nov 10, 2019 - Jan 5, 2020
(closed for review but you can still tweet)

NOTE: This is an unreviewed Preprint

Warning: This is a unreviewed preprint (What is a preprint?). Readers are warned that the document has not been peer-reviewed by expert/patient reviewers or an academic editor, may contain misleading claims, and is likely to undergo changes before final publication, if accepted, or may have been rejected/withdrawn (a note "no longer under consideration" will appear above).

Peer-review me: Readers with interest and expertise are encouraged to sign up as peer-reviewer, if the paper is within an open peer-review period (in this case, a "Peer-Review Me" button to sign up as reviewer is displayed above). All preprints currently open for review are listed here. Outside of the formal open peer-review period we encourage you to tweet about the preprint.

Citation: Please cite this preprint only for review purposes or for grant applications and CVs (if you are the author).

Final version: If our system detects a final peer-reviewed "version of record" (VoR) published in any journal, a link to that VoR will appear below. Readers are then encourage to cite the VoR instead of this preprint.

Settings: If you are the author, you can login and change the preprint display settings, but the preprint URL/DOI is supposed to be stable and citable, so it should not be removed once posted.

Submit: To post your own preprint, simply submit to any JMIR journal, and choose the appropriate settings to expose your submitted version as preprint.

Warning: This is an author submission that is not peer-reviewed or edited. Preprints - unless they show as "accepted" - should not be relied on to guide clinical practice or health-related behavior and should not be reported in news media as established information.

Intelligent alarms using principal component analysis on hemodynamic monitoring

  • Silje Skeide Fuglerud; 
  • Mikael Dyb Wedeld; 
  • Harald Martens; 
  • Nils Kristian Skjærvold; 

ABSTRACT

Background:

Patient monitors in modern hospitals give heartbeat waveform data that is reduced to aggregated variables and simple thresholds for alarms. Often, the monitors give a steady stream of non-specific alarms, leading to alarm fatigue in clinicians. An alarm can be seen as a classification problem, and by applying Principal Component Analysis (PCA) to the heart rate waveform of readily available monitoring devices, the accuracy of the classification of abnormality could be highly increased.

Objective:

To investigate whether physiological changes could be detected by looking at the heart rate waveform.

Methods:

A dataset of a healthy volunteer monitored with electrocardiography (ECG) and invasive blood pressure (BP) experiencing several tilts on a tilting table was investigated. A novel way of splitting continuous data based on the heartbeat was introduced. PCA was applied to classify the heartbeats.

Results:

A classification using only the aggregated variables heart rate (HR) and BP was able to correctly identify 20.7% of the heartbeats in the vertical tilt as abnormal. A classification using the full waveforms and combining the ECG and BP signals was able to correctly identify 83.5% of the heartbeats in the vertical tilt as abnormal. A humanistic machine learning (ML) method is then proposed based on the PCA classification.

Conclusions:

A ML method for classification of physiological variability was described. The main novelty lies in the splitting of an ECG and BP signal by the heart rate and performing a PCA on the data-table.


 Citation

Please cite as:

Fuglerud SS, Wedeld MD, Martens H, Skjærvold NK

Intelligent alarms using principal component analysis on hemodynamic monitoring

JMIR Preprints. 12/11/2019:16997

DOI: 10.2196/preprints.16997

URL: https://preprints.jmir.org/preprint/16997

Download PDF


Request queued. Please wait while the file is being generated. It may take some time.

© The authors. All rights reserved. This is a privileged document currently under peer-review/community review (or an accepted/rejected manuscript). Authors have provided JMIR Publications with an exclusive license to publish this preprint on it's website for review and ahead-of-print citation purposes only. While the final peer-reviewed paper may be licensed under a cc-by license on publication, at this stage authors and publisher expressively prohibit redistribution of this draft paper other than for review purposes.